Tubing Set Instructions for Use
1. Press BOTH buttons on the Tubing Set connector to connect & disconnect.
2. Connect the Tubing Set to the Fluid Connector located on the Top Assembly and
VitalWrap.
Troubleshooting
Frequently asked questions in troubleshooting the RecoveryWrap
System.
1. What do I do if the VitalWrap does not seem to get cold?
A) Check to make sure that the Fluid Canister was properly filled with ice
and water.
B) Check that the Tubing Set is properly connected and latched.
C) Check that fluid is flowing through the System and not obstructed.
D) Check that the temperature adjustment knob is set fully clockwise
(largest dot).
E) Check that the RED dip tube caps are removed prior to operation.
F) Check that power is applied.
G) Refer to Service if these actions are ineffective.
2. What do I do if the Tubing Set does not connect to the VitalWrap
or control unit?
A) Check to make sure that the indictor dots are aligned.
B) Check to make sure no foreign objects block the fluid connector.
C) Refer to Service if these actions are ineffective.
3. What do I do if the system does not turn on?
A) Check power connection.
B) Check that the Tubing Set is properly connected and latched.
C) Check that fluid is flowing through the System and not obstructed.
D) Wait 20 minutes and cycle power.
E) Refer to Service if these actions are ineffective.
4. What do I do if the system is not pumping any fluid through the
wrap?
A) Check that the Tubing Set and Vital Wrap are connected properly.
B) Or purge the uid pump by rst disconnecting the Tubing Set from the
RecoveryWrap System. With the RecoveryWrap System POWERED and turned
ON, press the Pump Troubleshooting Tool into the left connector until water
is pumped into the tool canister. Empty contents when canister becomes full
by removing cap. Repeat for approximately 2 minutes (see diagram below).
Reconnect the Tubing Set to the System and continue use.
C) Refer to Service if these actions are ineffective.